dc.description |
Tese (doutorado) - Universidade Federal de Santa Catarina, Centro de Ciências da Saúde, Programa de Pós-Graduação em Ciências Médicas, Florianópolis, 2020. |
|
dc.description.abstract |
Este estudo propõe uma nova metodologia para investigar o potencial de desconexão (Endoleak Tipo III) de pares de endopróteses sobrepostas em um modelo de vaso poplíteo submetidos a um teste de carga cíclica comparando três diferentes comprimentos de sobreposição. Um teste de fadiga cíclica foi desenvolvido para simular as cargas fisiológicas e movimentos aos quais a artéria é submetida durante a marcha (extensão, flexão, torção, compressão extrínseca e distensão diametral). O desenvolvimento do experimento foi baseado nas normas técnicas ASTM F2477-07 e ASTM F2942-13. Migração e desconexão foram medidos por DIC (Digital Image Correlation) para três comprimentos de sobreposição (20, 30 ou 40 milímetros). A metodologia desenvolvida neste estudo foi eficiente em criar um ambiente simulado que avalie a influência da marcha em endopróteses sobrepostas. Os resultados obtidos demonstram migração relevante mínima ou ausente entre as endopróteses (variação -0,06 a 0,34 milímetros). Conclui-se que a metodologia proposta é uma ferramenta importante para investigar a influência da biomecânica aos quais os dispositivos são submetidos sobre migração de endopróteses sobrepostas.<br> |
|
dc.description.abstract |
Abstract: This work proposes a new methodology to investigate the potential for disconnection (Type III endoleak) of pairs of overlapped endoprosthesis in a popliteal model vessel after a cyclic physiologic load, for three different overlap lengths. A multiaxial fatigue testing was designed to mimic the physiological loads and movements to which the arteries are submitted during gait (extension, flexion, torsion, extrinsic compression and diametral distension). The experiment design was based on the technical standards ASTM F2477-07 and ASTM F2942- 13. Migration and disconnection were monitored by DIC (Digital Image Correlation) for three different overlap lengths (20, 30 or 40mm). The methodology developed in this work was efficient to provide a simulated environment to evaluate the influence of gait biomechanics on overlapped endoprosthesis disconnection. Obtained results demonstrated minimal or absence of relevant migration between the endoprosthesis, range -0.06 to 0,34 millimeters. The proposed methodology was verified as a valuable tool to investigate the influence of the biomechanical environment which the devices are submitted to on the migration of overlapped endoprosthesis. It may become a new alternative to study the pre-clinical in vitro performance of single endoprosthesis or multiple connected devices with different overlapped regions |
en |
dc.format.extent |
63 p.| tabs., gráfs., fots. |
|
dc.language.iso |
por |
|
dc.subject.classification |
Ciências médicas |
|
dc.subject.classification |
Artéria poplítea |
|
dc.subject.classification |
Aneurisma |
|
dc.subject.classification |
Procedimentos endovasculares |
|
dc.title |
Avaliação in vitro das conexões dos dispositivos utilizados na correção do aneurisma de artéria poplítea |
|
dc.type |
Tese (Doutorado) |
|
dc.contributor.advisor-co |
Roesler, Carlos Rodrigo de Mello |
|