Projeto de fonte de corrente senoidal para medidas de sinal de bioimpedância
Show simple item record
dc.contributor |
Universidade Federal de Santa Catarina. |
pt_BR |
dc.contributor.advisor |
Schneider, Márcio Cherem |
|
dc.contributor.author |
Paes Neto, Jucemar |
|
dc.date.accessioned |
2024-01-29T17:14:02Z |
|
dc.date.available |
2024-01-29T17:14:02Z |
|
dc.date.issued |
2023-12-14 |
|
dc.identifier.uri |
https://repositorio.ufsc.br/handle/123456789/254128 |
|
dc.description |
TCC (graduação) - Universidade Federal de Santa Catarina, Centro Tecnológico, Engenharia Eletrônica. |
pt_BR |
dc.description.abstract |
Equipamentos de medida para aplicações médicas não invasivas se tornam cada vez
mais comuns para realizar diagnósticos de patologias e a integração de wearables no
cotidiano das pessoas se torna cada vez mais comum. Além disso, análises por
processamento de imagem e a disseminação de tecnologias de machine learning tem
tornado técnicas como a Tomografia de Impedância Elétrica alternativas viáveis para
diagnósticos de lesões ortopédicas e até mesmo ensaios em tempo real para atletismo
de alto desempenho. Contudo, para atingir tais objetivos de dispositivos altamente não
intrusivos e que realizem medidas confiáveis são necessários um alto gral de
integração de componentes eletrônicos e técnicas adequadas de instrumentação. Por
isso, esse trabalho propõe-se a desenvolver um driver de corrente para a realização
de medidas de bioimpedância através do método de Síntese Digital Direta. Isso
porque essa técnica possui diversas vantagens sobre suas contrapartes analógicas,
como redução na dependência da temperatura, controle superior sobre a frequência
de saída e ajuste fino sobre a fase através de controle digital além de permitir alto grau
de integração por não requerer componentes reativos dispendiosos em área.
Utilizando essa técnica e através de otimizações das lógicas e circuitos digitais foi
obtido um circuito capaz de fornecer uma saída com erro de frequência de 5,95 ppm
com clock máximo de 512,5 MHz e distorção harmônica total de 0,258%. |
pt_BR |
dc.description.abstract |
Measurement equipment for non-invasive medical applications is becoming
increasingly common to diagnose pathologies and the integration of wearables into
people's daily lives is becoming more and more common. In addition, image
processing analyzes, and the spread of machine learning technologies have made
techniques such as Electrical Impedance Tomography viable alternatives for
diagnosing orthopedic injuries and even real-time tests for high-performance athletics.
However, to achieve such goals of highly non-intrusive devices that perform reliable
measurements, a high degree of integration of electronic components and adequate
instrumentation techniques are required. Therefore, this work proposes to develop a
current driver to carry out bioimpedance measurements through the Direct Digital
Synthesis method. This is because this technique has several advantages over its
analog counterparts, such as reduced temperature dependence, superior control over
the output frequency and fine adjustment over the phase through digital control, in
addition to allowing a high degree of integration by not requiring area consuming
reactive components. Using this technique and trough logical and digital circuitry
optimizations the circuit achieved was able to supply an output with 5,95 ppm
frequency error with a maximum clock of 512,5 MHz and total harmonic distortion of
0,258%. |
pt_BR |
dc.format.extent |
72 f. |
pt_BR |
dc.language.iso |
pt_BR |
pt_BR |
dc.publisher |
Florianópolis, SC. |
pt_BR |
dc.rights |
Open Access. |
en |
dc.subject |
Bioimpedância |
pt_BR |
dc.subject |
Síntese Digital Direta |
pt_BR |
dc.subject |
Driver de Corrente |
pt_BR |
dc.title |
Projeto de fonte de corrente senoidal para medidas de sinal de bioimpedância |
pt_BR |
dc.type |
TCCgrad |
pt_BR |
Files in this item
This item appears in the following Collection(s)
Show simple item record
Search DSpace
Browse
-
All of DSpace
-
This Collection
My Account
Statistics
Compartilhar